Content Status
Type
Linked Node
CBNAAT Hardware Problem: Modules Not Detected
Learning ObjectivesCBNAAT Hardware Problem: Modules Not Detected
H5Content
Content
One of the common problems encountered while using the CBNAAT machine is the non-detection of the module.
Usually, there is a loss of communication between modules and software, leading to non-detection.
Origins:
- Ethernet connection between Personal Computer (PC) and CBNAAT is bad
- Power supply issue (main power or Universal Power Supply (UPS) fluctuations)
- Bad connection points between gateway board and modules
- Too high room temperature
Solutions:
- Restart the instrument first and then the computer.
- Unplug and re-plug the ethernet cable between the PC and the instrument, then restart as above.
- Secure the power supply and use an adapted UPS/ surge protector.
- Check the temperature inside the module (should be below 39°C).
Figure: Troubleshooting protocol for the non-detection of CBNAAT Module
Video file
Video : CBNAAT Hardware Problem: Modules Not Detected
Resources
LMS Page Link
Content Creator
Reviewer
Target Audience
- Log in to post comments