Under the National TB Elimination Program (NTEP), each Cartridge-based Nucleic Acid Test (CBNAAT) site needs to keep a maintenance record for the instrument.

 

Cleaning and disinfecting the system components is crucial for proper CBNAAT maintenance. The tasks and frequency are shown in the table below.

 

Table: Maintenance of CBNAAT - tasks and frequency
TASK FREQUENCY
Clean work area Daily
Close all module doors Daily
Discard used cartridges Daily
Power down the CBNAAT instrument Weekly
Power down the CBNAAT computer Weekly
Archive tests Monthly
Purge tests Monthly
Replace the fan filters Monthly
Clean plunger rod and cartridge bay Quarterly
Clean the instrument surfaces Quarterly
Perform annual instrument maintenance Annually
Print system log report As necessary
Back up database As necessary

 

For routine surface cleaning 70% ethanol is used; while a final concentration of 1:10 dilution of Sodium hypochlorite is used for disinfection of different parts of the CBNAAT instrument followed by removing bleach residue with 70% ethanol.

 

The logs for the maintenance of records for CBNAAT is shown in Figure 1.

 

Figure 1: Maintenance Log Sheet

 

 

The maintenance log sheet is also provided by the vendor as shown in Figure 2. The monthly log for maintenance records can be obtained from the CBNAAT System Operator Manual CDROM that can be copied and used.

 

Figure 2: Maintenance Log Sheet provided by the Vendor

 

 