Here we discuss the error classes and actions in Truelab.

 

Truelab Errors - Classes and Actions

Error Class

Error Name

Reason for the error

Error Actions

Error 1

Thermal Cycling Error

Occurs when the test chip is faulty and the thermal cycling does not happen.

Re-run the same elute using another chip. If the error repeats, process the sample again and re-run the elute using another chip. If the problem still persists, contact the Molbio support team.

Error 2

Test Stopped Manually

Occurs when the user has manually stopped the ongoing test and the analyser does not have sufficient run time to compute data.

Error 3

Incorrect Optical Profile

Occurs when there is a deviation in the expected optical profile due to a reduction in reaction volume in the chip during the course of the reaction.

Error 4

Runtime Error

Occurs when run data capture/ analysis is incomplete.

Error 5

Probe Check Error

Occurs in the event of a low initial signal due to insufficient mastermix dispensed onto the chip.

Invalid

 

Whenever Truelab displays the result as 'INVALID', it means that the internal control did not amplify in Polymerase Chain Reaction (PCR) or sample extraction was not proper.

 

Resources

 

Assessment

Question    

Answer 1    

Answer 2    

Answer 3    

Answer 4    

Correct answer    

Correct explanation    

Page id    

Part of Pre-test    

Part of Post-test    

When does thermal cycling error occur in Truelab's real-time micro PCR analyzer?

Occurs when the test chip is faulty and the thermal cycling does not happen.

Occurs when run data capture/ analysis is incomplete.

Occurs in the event of low initial signal.

Occurs when the user has manually stopped the ongoing test.

    1

Thermal cycling error occurs when the test chip is faulty and the thermal cycling does not happen.

    

   Yes

 Yes

When does the Runtime Error occur in Truelab's real-time micro PCR analyzer?

Occurs when the test chip is faulty and the thermal cycling does not happen.

Occurs when run data capture/ analysis is incomplete.

Occurs in the event of low initial signal.

Occurs when the user has manually stopped the ongoing test.

      2 Runtime error occurs when run data capture/ analysis is incomplete.  

 