CBNAAT Troubleshooting: Error- 2008 and 2009
Learning ObjectivesCBNAAT Troubleshooting: Error- 2008 and 2009
H5Content
Content
During the use of the Cartridge-based Nucleic Acid Amplification Test (CBNAAT) machine, Error- 2008 and 2009 may appear on the screen and needs troubleshooting.
Problem:
2008: Pressure reading exceeds the maximum
2009: Syringe pressure is below the protocol limit
Figure: Error 2008
Origins:
- The cartridge filter is clogged (due to too viscous sample or particles)
- Pressure sensor failed
Solutions:
- Make sure the sample does not contain any solid particles.
- Make sure the sample is totally liquified before transferring it to the cartridges.
- If after 15 minutes of incubation with the sample reagent, the sample is still too viscous, wait up to 10 more minutes.
- Use a new cartridge and add only a sample reagent to test. If the problem persists, it is likely a module problem: Contact Technical Support.