MGIT 960 Instrument: Maintenance
Learning ObjectivesLearn about the three types of maintenance needs and specific tasks under each to ensure the proper functioning and reliability of the MGIT 960 instrument.
H5Content
Content
The Mycobacterium Growth Inhibitor Tube (MGIT) instrument needs daily, routine and need-based maintenance.
Maintenance Needs
- Daily
- Checking the incubator temperature
- Checking the temperature of the internal thermometer
- Checking the printer paper supply
- Testing the red, green and amber Light-emitting Diode (LED) lights for each drawer
- Printing the "Quality Control Report”
- Routine, user preventive
- Monthly changing or cleaning of the cabinet air filters
- Replacement of the fluorescence calibrator vials when they expire
- “As needed” basis
- Cleaning the Barcode Scanner Window with a damp, lint-free, non-abrasive cloth
- Cleaning of any leakage or spill of drawer tube holder with disinfectants (1% sodium hypochlorite, 70% alcohol)
- Maintenance/ repair by the manufacturer
Maintenance and cleaning logs should be maintained for the instrument.