Content Status

Type

Linked Node

H5Content
Content

In the Line Probe Assay (LPA) lab, the water bath is used for warming reagents used in the Hybridization Room.

It is electrically heated and electronically controlled.

 

Components:

  • Recessing push/ turn control for simple operation of the bath
  • Visual temperature indication​
  • Push button to set desired temperature​
  • Electrically heated and electronically controlled​
  • Water trough​

 

Installation of Water Bath (Figure):

  • The unit must be placed on a horizontal, non-flammable sturdy surface. ​​
  • The vent openings must remain unobstructed. ​​
  • Minimum wall spacing on all sides is 80 millimetres. ​​
  • The minimum spacing from top of the bath to the next ceiling must be 750 millimetres. ​

Figure: Spacing during Installation of Water Bath (Front and Side View)

 

 

Working with the Water Bath: ​

The heating positioned on three sides around the tank ensures a natural water circulation of the liquid inside, thus securing an optimal uniform temperature distribution​.

  • Vent openings must be unobstructed​.
  • The surface must be horizontal, nonflammable during installation​.

 

General Instructions for Maintenance

  • Always switch off the appliance before any maintenance activity​.
  • Always record the maintenance activity performed​.
  • Rusting objects must not come into contact with stainless steel bath tank​.
  • Rusting areas must be cleaned and polished immediately​.
  • Lubricate/ oil the hinge bolts from time to time​.
  • Replace distilled/ demineralized water and clean the water bath every week​.

 

Annual Maintenance

  • This is done by a qualified service technician
  • Includes the calibration or validation of the temperature with a thermostat/ digital controller with devices traceable to National/ International Standards​.

 

Resources

 

 

Kindly provide your valuable feedback on the page to the link provided HERE
 

Content Creator

Reviewer