Content Status

Type

Linked Node

H5Content
Content

Daily maintenance of Truelab and Trueprep instruments performed by the laboratory personnel includes:

  • Maintaining a dust-free environment by wiping the exterior of instruments with a dry, lint-free cloth
  • Cleaning the surface of the Trueprep device once with 70% ethanol
  • Cleaning the spill or leak with a cloth or tissue-paper dipped in disinfectant (1% sodium hypochlorite solution, followed by 70% ethanol)
  • Not spilling water or any other solution on the surface of instruments
  • Discarding used cartridges, chips, reagent bottles and other consumables in 1% sodium hypochlorite solution, soaking for minimum 30 minutes, and disposing as per the biomedical waste management guidelines.

 

Monthly maintenance of the Truelab Analyzer performed by laboratory personnel includes:

  • Cleaning the surface of the Truelab Analyzer with 70% ethanol
  • Cleaning the Truelab bays
  • Calibrating temperature.

 

Annual/ need-based maintenance of the Truenat instrument performed by the laboratory personnel/ manufacturer includes:

  • Flush protocol for the Trueprep device when the device is left idle (not used) for 10 days and/ or errors relating to extraction process occur
  • Replacement of spillage tray or linear motion guide tray for the Trueprep device in case of sample spillage on trays during extraction or when cross-contamination is suspected
  • Replacement of slider glass for Truelab Analyzer after 200 tests and/ or when related errors occur
  • Temperature calibration for Truelab Analyzer when error related to temperature occur and/ or when the temperature curve is abnormal (shows blips)
  • Calibration of micropipettes (every 6 months/ biannual).

 

Resources

 

 

Kindly provide your valuable feedback on the page to the link provided HERE

Content Creator

Reviewer

Target Audience