Annual Maintenance of the Truenat Instrument
Learning ObjectivesAnnual Maintenance of the Truenat Instrument
H5Content
Content
Need-based maintenance for Truenat instrument is done by lab personnel/ manufacturer while the manufacturer does annual maintenance.
The tasks included under annual maintenance are:
- Flush protocol for Trueprep device when the device is left idle (not used) for 10 days/ annually and/or errors relating to extraction process occur.
- Spillage tray or linear motion guide tray replacement for Trueprep device in case of sample spillage on trays during extraction or when cross-contamination is suspected/ done annually.
- Slider glass replacement for Truelab Analyzer after 200 tests/ annually and/or when related errors occur.
- Temperature calibration for Truelab Analyzer when error related to temperature occur and/or when temperature curve is abnormal (shows blips)/ annually.
- Biannual calibration/replacement of micropipettes.
Resources
- Practical Guide to Implementation of Truenatâ„¢ Tests for the Detection of TB and Rifampicin Resistance.
- Trueprep Manual.