Unloaded MGIT DST Report
Learning Objectivesdescribe steps
Unloaded MGIT DST Report
A completed DST in the Bactec MGIT-960 instrument is indicated by a red “+” on the drawer. The following steps are involved in removing the completed DSTs from the instrument and printing the “Unload AST Set” report:
Step 1: Open the desired drawer. Press the “Remove Completed AST Sets”
Step 2: The first completed AST set stations illuminate with FLASHING GREEN indicators.
Step 3: Remove the carrier, starting with the completed set closest to the front of the drawer, and scan its barcode label. The LEDs at this station extinguish.
Step 4: Repeat steps 2 – 3 to remove additional AST sets.
Step 5: Place completed AST sets in the AST tube rack.
Step 6: Close the drawer and press the “printer” soft key to access the report selection.
Step 7: Press the “Unloaded AST set report” soft key to print the report.
Note: Ensure that a working printer is connected to the instrument and check that the paper is present before printing the report.