Content Status

Type

Linked Node

H5Content
Content

The beauty of the MGIT 960 tubes is that they can be read by the instrument and also manually. There are times AFB cultures are incubated at temperatures other than 37oC. If that is the case, the MGIT tube can be incubated at an alternate temperature. These tubes would then need to be read manually with a Wood’s Lamp or other UV sources of the proper wavelength and intensity.

In the event of an instrument failure where the instrument is down for an extended period, you may decide to incubate the tubes offline. These tubes can also be read manually.

Manual readings should not be conducted in a darkened room or rooms with bright sunlight. Normal room lighting is appropriate. When using any UV source, one must always wear protective UV goggles.

To read the tubes manually, place the rack of tubes directly on top of the UV lamp. There is no need to take the tubes out of the rack. Some accounts leave a blank row between tubes. This can make it easier to see the fluorescence when reading the tubes.

Tilt the tubes/rack slightly so that the tubes are read at a slight angle. Look for fluorescence both at the sensor at the bottom of the tube and the meniscus at the top of the media line (the fluorescence of the sensor reflects in the meniscus). Place a positive and negative control at the end of the row to give visual assistance in the reading. Place the suspect between these two tubes to determine if the patient tube is positive. The fluorescence in the patient tube should be greater than the background fluorescence in the negative control but does not need to be as great as the fluorescence in the positive control to be considered positive.

Resource

Assessment

Question 1

Answer 1

Answer 2

Answer 3

Answer 4

Correct Answer

Correct Explanation

Page id

Part of Pre-Test

Part of Post-Test

The MGIT tubes can be read manually.

 

True

False

 

 

True

The beauty of the MGIT 960 tubes is that they can be read by the instrument as well as manually.

 

Yes

Yes

Question 2

 

 

 

 

 

 

 

 

 

UV sources with proper wavelength and intensity can be used to read MGIT tubes.

True

False

 

 

True

 

Wood’s Lamp or other UV sources of the proper wavelength and intensity can be used to read MGIT tubes manually.

 

Yes

Yes

 

Content Creator

Reviewer

Target Audience